Chapter 9
Page no. 1272
JC-DR-A-232.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A232 - Lift Screw Assembly
pression Arm to the top of the Gantry.
If you secured the Compression Arm with the Gantry LOTO tool, perform the following sub-steps:
a. Push the knob (6) upwards until its maximum height.
b. Gently pull the lower part of the Gantry LOTO Tool away from the Gantry framework.
c. Fully remove the Gantry LOTO Tool paying particular attention to not drop it.
CAUTION
The Gantry LOTO Tool is 0.7 kg. Be prepared to take the weight of the Gantry LOTO Tool
as you pull it s lower apart away from the Gantry. If you are not prepared to take the weight
of the Gantry LOTO Tool, there is a risk that it could fall on the lower parts of the Gantry
(e.g. the Lift board) or your feet.
8. Adjust the lift position by rotating the Lift Screw anti-clockwise (7) (13 mm or 17 mm hexagonal
6