Chapter 9
Page no. 1176
JC-DR-A-204.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A204 - Lift Potentiometer
6
PROCEDURE
6-1
Disassembly of the Lift Potentiometer
1. Release the four bolts securing the lower X plate (10 mm open-ended wrench).
2. Disconnect from the lift board (2) the Cable W203: Potentiometer cable (connector J11).
3. Release without removing the four screws (3) secur-
ing the potentiometer (2.5 mm allen wrench).
4. Slide the potentiometer out of its supports.
2
J11
3