GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 893
Chapter 9
ERR-SUB-A-003.fm
132/000
INHIBIT
TILT INVALID PDO
LENGTH
Change the Tilt board then perform a Tilt Calibration. D/R A266
CAL A027
132/001
INHIBIT
Tube Tilt software
error
The software/firmware on the Tube Tilt board is cor-
rupt.
This failure is linked to a firmware problem (micro-
controller or CPLD). Change the Tilt board then per-
form a Tilt Calibration.
D/R A266
CAL A027
132/002
WARNING
Tube Tilt fan error
The rotation speed of the FAN is monitored through
a third wire. If it is not within an acceptable range,
this signal is produced
Change the Fan and cable W406.
132/003
HALT
Tube Tilt switch
defect
SwitchDefect [i] (for i = 1 - 9) occurs when the two
signals from the button do not arrive at the same
time on the CPLD.
The error signal indicates the nature of the button but
not the side right or left.
- Lift up (V1, V2)
- Lift down (V1, V2)
- Rotation Right (V1, V2)
- Rotation Left (V1, V2)
- Light/Field Of View
- MLO +/-
- CC
Refer to
JC DIAG A1301 CONTROL KEYPAD FAIL-
to fix the problem.
A1301
132/004
HALT
Tube Tilt switch
defect
132/005
HALT
Tube Tilt switch
defect
132/006
HALT
Tube Tilt switch
defect
132/007
HALT
Tube Tilt switch
defect
132/008
HALT
Tube Tilt switch
defect
132/009
HALT
Tube Tilt switch
defect
132/010
HALT
Tube Tilt switch
defect
132/025
HALT
Tube Tilt switch
defect
132/026
HALT
Tube Tilt switch
defect
132/027
HALT
Tube Tilt switch
defect
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card