GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A410 - Inverter Board 300-PL1
Page no. 1597
Chapter 9
JC-DR-A-410.fm
6-2
Installation of the Inverter Board 300-PL1
1. Carefully place the new Inverter board on the Generator chassis. Ensure that the one thread passes
thorough the Inverter board top middle hole, and ensure that the holes for the six bolts line up with
the four threads on the Generator chassis.
2. When placing the new Inverter board on the Generator chassis ensure that you the thread the three
sets of brown cables through the holes in the Inverter board (1, 2, 3).
3. Attach the bolt (4) (7 mm open ended wrench) to the thread.
4. Attach the six bolts (5) (7 mm allen wrench) to secure the rest of the Inverter board to the Generator
chassis.
5. Connect all the cables to the new Inverter board as summarized in the diagram and table below.
Connec-
tor Label
Tool Description
of
cable
(usually identified by code on a white label)
Cable
color
XJ1
N/A
300 PL1 XJ1 2110262
(red line at top of ribbon connector)
Gray
X1
10 mm open ended wrench
X1
Black
X2
N/A
Empty
N/A
X3
10 mm open ended wrench
X3
Brown
X4
10 mm open ended wrench
X4
Brown
X5
10 mm open ended wrench
X5
Black
X6
10 mm open ended wrench
X6
Brown
X7
10 mm open ended wrench
X7
Brown
X8
10 mm open ended wrench
Two cables, one from below and one from behind: X8
Brown
X9
10 mm open ended wrench
X9
Brown
5
4
1
2
3