GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
Page no. 1025
Chapter 9
JC-TSG-A-010.fm
ILLUSTRATION 3 - STEPS TO PERFORM IF BAD PIXELS EXIST AT DETECTOR EDGES
Bad Pixel calibration passes?
Perform collimator calibration
Job Card CAL A029 - Collimator Format Calibration
YES
NO
1.
Adjust the Collimator positioning.
2.
Perform collimator calibration
Job Card CAL A029 - Collimator Format Calibra-
3.
Perform Bad Pixel Calibration
Job Card CAL A041 - Bad Pixel Calibration
Perform post-corrective actions
(see
,
be performed after bad pixel
calibration
Perform Bad Pixel Calibration
Job Card CAL A041 - Bad Pixel Calibration
Bad Pixel calibration passes?
YES
NO
1.
Change the Collimator
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
2.
Perform collimator calibration
(see
Job Card CAL A029 - Collimator Format Calibration
)
3.
Perform Bad Pixel Calibration
(see
Job Card CAL A041 - Bad Pixel Calibration
Bad Pixel calibration passes?
The customer must not use the system.
Use the service escalation process.
YES
Perform post-corrective actions
(see
,
performed after bad pixel calibration
).
NO
Perform post-corrective actions (see