Page no. 382
JC-LOTO-A-001.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card LOTO A001 - Lock-Out / Tag-Out (LOTO) Procedures
8-5
Apply LOTO
You must apply LOTO to the power source of the Line Supply Cable (i.e. the Mains Distribution Panel).
1. If the activity requires one Field Engineer, install the red LOTO padlock onto the Mains Distribution
Panel, as follows:
•
On the Mains Distribution Panel insert the red LOTO padlock and red LOTO label to isolate
incoming supply.
•
If there exists a connection between the socket on the Mains Distribution Panel and the entrance
into the Generator, then secure the yellow box with an additional red LOTO padlock and red
LOTO label.
•
The key and the red LOTO padlock is the personal property of the individual Feild Engineer and
must never be loaned to others.
If the activity requires more than one Field Engineer install the Multihasp onto the Mains Distribution
Panel:
•
If required, use the appropriate RED plastic adapter from the LOTO kit to lock-out individual
breakers according to size or type.
•
Use a multihasp in conjunction with a red LOTO padlock and LOTO red tag for each Field Engi-