GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
Page no. 651
Chapter 7
JC-ELE-A-021.fm
18. At the
Offset from GMT
screen that appears,
type
0
for the offset the press
<Return>
to confirm the
offset value. Press the
<F2>
key to continue.
19. A
Date and Time
screen appear with some existing date and time settings.
Proceed as follows:
•
If the date and time settings are incorrect, use the Up or Down arrow keys to move to
the field to
be modified (as applicable). Type in the correct value, then press
<Return>
to confirm the new
value and move to the next field. When you are happy with the date and time settings, press the
<F2>
key to continue.
•
If the date and time settings are correct, press the
<F2>
key to continue.
Specify the number of hours difference between ....
0
Offset From GMT
Year (4 digits):
> Accept the default date and time or enter
new values.
Date and Time
2007
Month (1-12): 06
Day (1-31): 09
Hour (0-23): 11
Minute: 33
Date and time: 2007-06-09 11:33