ERR SUB A003 - Gantry Error Messages
Page no. 925
Chapter 9
ERR-SUB-A-003.fm
166/006
INHIBIT
TCP/IP PACKET
WITH WRONG SIZE
The Gantry received a TCP/IP packet originating
from another node in the Senographe System, but
rejected it because it was the wrong size.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
166/007
INHIBIT
ADS IS NOT CON-
NECTED - TRANS-
MIT FAILED
A message from the Gantry CPU can not be sent to
ADS because the ADS is not connected or powered
down.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
167/001
HALT
PDU NODE NOT
STARTED
The Power Distribution Unit (PDU) node did not start.
At power-up, the PDU node sends an acknowledg-
ment (which can be different from the first heartbeat
signal) to Gantry CPU. If the PDU node does not
send an acknowledgment, the Gantry CPU creates
this error.
1. Reboot the system from the X-ray Console and
perform a compression motion before a com-
plete Gantry boot.
2. If the problem persists, change the PDU Board
(Firmware or Hardware problem) (see
3. If the problem persists, check CAN bus.
167/002
HALT
PDU NODE NOT
RESPONDING
The Power Distribution Unit (PDU) node did not send
a heartbeat signal.
Between two motions, the PDU node sends a Heart-
beat signal every second to Gantry CPU. If not, the
Gantry CPU creates this error.
During lift motion, the lift PDU sends an actual value
message regularly in addition to the heartbeat. If not,
the Gantry CPU creates this error.
Possible FRUs= PDU board, CAN bus
167/003
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown error code
from the Power Distribution Unit (PDU) node.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card