manualshive.com logo in svg
background image

Содержание Drying Cabinet

Страница 1: ......

Страница 2: ...of Contents Forward 3 Design Parameters 3 Quality Policy 4 Section A Unit Operation 5 Section B Unit Maintenance 6 Trouble Shooting Guide 7 Wiring Diagram 8 Section C Unit Installation 9 Unit Specific...

Страница 3: ...r tubing face masks anaesthetic bags and other ancillary equipment A total of six 6 stainless steel shelves is provided The shelves are removable with 6 positions including the base The cabinet is ins...

Страница 4: ......

Страница 5: ...net appears for approximately five seconds After five seconds the unit switches to display the actual cabinet temperature Changing the set temperature lock out feature AUTHORISED PERSONNEL ONLY Author...

Страница 6: ...m water and detergent Wipe over outside panels with stainless steel cleaner Weekly Remove filter on air intake top right hand side panel and wash in warm water and replace when dry Half Yearly Check e...

Страница 7: ...turns off Fan failure Switch off power supply at main and Call for Service Display is on but unit not heating Element failure Relay board failure Switch off power supply at main and Call for Service...

Страница 8: ...Drying Cabinet Operation Maintenance and Installation Manual Issue 15 Page 8 7th October 2011 Wiring Diagram...

Страница 9: ...be a 5 to10mm gap between the side panel and the timber Clearance on the top of the Drying Cabinet needs to be typically 500mm This allows access to replace the electric motor heating elements etc All...

Страница 10: ...B SKCM325 Access via compatible RS232 Control PCB Overtemp Protection Electronic cutout 5 C above set point Temperature Control Incremental 1 C 70 C Factory set at 70 C Secondary Overtemp Protection H...

Страница 11: ...ltered Parts damaged in transit back to Malmet Leeton due to poor packaging could result in warranty claim being rejected Any part tampered with or has unauthorised repairs and modifications will be r...

Страница 12: ...tial loss either direct or indirect resulting from power spikes incorrect operation incorrect installation or faulty product are excluded All warranties are NOT transferable Unauthorised use of non ge...

Страница 13: ...installer is responsible for the correct installation start up and demonstrating the operation of this product They are also responsible for issuing the relevant certificates of compliance These may...

Страница 14: ......