Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A006 - First System Power-On
Page no. 545
Chapter 7
JC-ELE-A-006.fm
7
POWERING DOWN THE SYSTEM
•
Normal Shutdown Procedure:
Switch power off at the AWS Monitor by clicking the
System Shutdown
button on the Browser win-
dow. Wait for complete system and UPS shut down (about 2 minutes); check that no LEDs remain lit
on the UPS before continuing.
•
To power down the Gantry and Generator system
only
, press the switch
STOP
on the X-ray Console.
•
Never switch off at the UPS directly except in emergency (risk of data loss).