Job Card PM A428 - Conditioner Coolant Level Check and Top-up
Chapter 11
1
SUPPLIES
•
2
TOOLS
•
•
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 5 minutes
4
SAFETY PRECAUTIONS
•
When filling or topping up the Conditioner, take great care to keep coolant away from the vents, the
ON/OFF switch and the power cord connection. Do not immerse the Conditioner in any liquid.
•
Dispose of old and surplus coolant with care as described in the section
Surplus Coolant Disposal
;
refer to the
Caution
below.
CAUTION
• Before adding coolant to the Conditioner, always switch OFF the Generator and MDR cir-
cuit breakers in the Generator Cabinet. Switch on the MDR breaker only as instructed.
• Always use a funnel when adding coolant to the Conditioner.
• Always wear protective gloves and goggles when working with coolant.
• When working with coolant you should treat it with care; avoid unnecessary contact .
See below for First Aid advice:
First Aid in case of contact with Conditioner coolant
:
Immediately flush eyes or skin with plenty of water for at least 15 minutes.
If coolant contacts the eye, call a physician immediately.
If coolant is swallowed, drink one to two glasses of water or milk. Avoid alcohol. Call a physician
immediately.
Remove and wash contaminated clothing before reuse.
5
PREREQUISITES
Remove the Generator Cabinet covers as required for access to the Conditioner. Refer to
A042 - Remove/Reinstall Generator Covers
6
PROCEDURE
CAUTION
When performing any of the following procedures, be careful to avoid spillage of coolant,
especially onto or near electrical connections. If any coolant is spilled, it must be cleaned
up immediately.