GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A411 - Mains Distribution Board 200-PL1
Page no. 1607
Chapter 9
JC-DR-A-411.fm
rotary switch in the Generator cabinet (position I).
5. Switch on the Senographe system electrical power (i.e. Generator, Gantry, and Control Station).
Wait until the Gantry boot is complete, and check that no error has been reported. If the Mains Distri-
bution board is operating correctly, the LED states on the Mains Distribution board are as follows:
•
red DE1 LED is OFF
•
green DE2 LED is ON
•
green DE3 LED is ON
•
red DE4 LED is OFF
•
yellow DE5 LED is OFF
•
yellow DE6 LED is ON
•
red DE7 LED is OFF
•
clear N1 LED is ON
•
clear K1 LED is ON
6. Reinstall the Generator covers (see
Job Card PHY A042 - Remove/Reinstall Generator Covers
7
COMPLETION
None.