GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A232 - Lift Screw Assembly
Page no. 1275
Chapter 9
JC-DR-A-232.fm
6-5
Grease the New Lift Screw
1. Apply Grease (for Lift and Rotation) (5212483) (
Orapi CT FLON 607
) to the Leadscrew by hand.
Ensure the Leadscrew is homogeneously covered with grease, as shown in illustration 1.
CAUTION
Do not use any other type of grease, other than Grease for Lift and Rotation (5212483)
(Orapi CT FLON 607).
Do not use a brush to apply the grease as this may introduce foreign objects on to the
Leadscrew. Only use your hands to apply the grease to the Leadscrew.
About 35 g (1.23 oz) of grease is sufficient to cover the entire length of the Leadscrew.