Chapter 7
Page no. 596
JC-ELE-A-020.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-2-2
Agfa 5500/5503 DICOM Printer
•
DICOM printer label: Any name you wish to see appearing on the Film Composer e.g., Drystar5503.
•
Hostname: Use printer Hostname if any, or a name unique to your Network e.g., drystar5503.
•
Network address: A unique IP address given by the hospital Network Administrator, e.g.,
3.48.16.136.
•
Application Entity Title (AET):
Default value is
DS5500
, however contact the Agfa engineer for
the correct AET value to use
•
Port number:
Must be
104
•
Layouts:
Must be
1 x 1
, 1 x 2
, 2 x 1
, or 2 x 2
•
Film size:
Must be
8 in x 10 in W = 3828 H = 4958
10 in x 12 in W = 4892 H = 5900
•
Pixel size
50
•
Configuration information:
Must be
PERCEPTION_LUT=OEM919
•
Density
Must be
Min: no value Max: 370
•
Magnification type:
Must be
Cubic
•
Smoothing factor:
Must be
143
•
Trim:
Must be
NO
•
Printer Memory Size
Must be
100