Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Page no. 729
Chapter 7
JC-ELE-A-043.fm
Job Card ELE A043 - AWS Monitor Calibration
Chapter 7
1
SUPPLIES
None
2
TOOLS
A calibrated Luminance Meter that permits measuring the luminance of the monitor in contact (Direct)
mode as illustrated below.
•
(P/N 2319223) must be used with the foam ring attached, and must be
configured in
Direct
mode.
•
The
Away/Direct
button toggles between away and direct modes. To select the
Direct
mode, keep
pressing the
Away/Direct
button until
D
is displayed on the lower left corner of the Spot Meter
screen.
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
Remove canon and place foam ring
Luminance Meter
Foam Ring
Direct
mode selected
Away/Direct
mode
button selector