GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Central Listing
Page no. 249
Chapter 3
Indicators and Switches.fm
2-2-4.
Generator Interface Board 400PL2 Test Points
Test Point
Label (in French)
Function (in English)
TP1
Cons_Chauf_A
Requested Heater A Current
TP2
Cons_Chauf_B
Requested Heater B Current (Ground)
TP3
T5000
5000 RPM Speed
TP4
Entretien
Normal Rotation Speed
TP5
Marche_Lanceur
Starter Working
TP6
Freq_Polar
Polar Frequency
TP7
kV_Vide
kV Empty
TP8
Pose_Rx
X-ray Exposure
TP9
Marche_Chauf
Heater Working
TP10
Piste_2
Track 2
TP11
mA_Max
Maximum mA
TP12
kV_INF_5_B
TP13
Strobe_kV_A
TP14
SRI_kV_A
TP15
Load_kV_A
TP16
Load_kV_B
TP17
Voyant_Pose_A
Exposure Warning A
TP18
0V_Pupitre
X-ray Console 0V
TP19
Button Prepare
TP20
mA_A
TP21
Rx +
TP22
Rx -
TP23
Tx -
TP24
Frein
Brake
TP25
Alim_Puis_OK
Power Supply OK
TP26
Secu_RatTrap
TP27
Pres_Supply
Supply Present
TP28
Pres_Chauf
Heater Present
TP29
Lect_Secu
TP30
Marche_Polar
Polarization Working
TP31
Pres_Polar
Polarization Present
TP32
Pres_Rotat
Rotation Present
TP33
En_Pose
On Exposure
TP34
Secu_Bloq
Blocking Security
TP35
Timeout_Regul
Timeout Regulation
TP36
Trad
TP37
mA_B
TP38
Strobe_kV_B