Chapter 3
Page no. 94
S2100D How To.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
In installation mode, the six function keys below the display window carry out functions or menu
selections which depend on the current procedure. The choice is indicated by a label displayed
above the key. The illustration below shows typical X-ray Console displays before (left) and after
(right) entering installation mode. The normal key symbols (shown shaded) no longer apply, and in
the example shown two of the keys (shown shaded) have no function. The remaining four carry out
the functions displayed above them (e.g., the left-hand key now gives access to the GENE menus).
•
Some special
installation menus
are displayed only after changing the position of the installation
enable switch (see below).
+0
SMALL AUTO
AUTO
A
SELECT ARM/GENE INSTALL
GENE
ARM
MEDICAL LANGUAGE