GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A315 - International Keyboard
Page no. 1445
Chapter 9
JC-DR-A-315.fm
7
PROCEDURE FOR V1/V2 CONTROL STATIONS
7-1
Determine the OpenBoot Version of the SB150 Workstation
1. From behind the Browser, launch a command window (see
Launching a Command Line Window
A command window appears with an ADS prompt.
2. Query the OpenBoot version of the SB150 Workstation, by entering:
prtconf -V
The OpenBoot version of the SB150 Workstation appears.
•
If the OpenBoot version is lower than 4.10.6, then you cannot change the keyboard with an Inter-
national Keyboard.
•
If the OpenBoot version is equal to or greater than 4.10.6, then shutdown the Control Station
using the shutdown button the Browser, and proceed to
to change the keyboard with
an International Keyboard.
WARNING
Before proceeding to
, shut down correctly the equipment and switch off the
Senographe system hospital supply circuit breaker. Lock it in the off position or mark it
clearly with a warning sign to prevent others from applying power
.
7-2
Connect the new keyboard
1. Unplug and remove the X-ray Console.
2. Remove Control Station Top Cover. See
Job Card PHY A040 - Remove/Reinstall Control Station
.
3. Remove Control Station Omega Cover. See
Job Card PHY A040 - Remove/Reinstall Control Station
.
4. Disconnect the original keyboard from the back of the ADS Computer, and remove it from the Control
Station.
4
3
Reference
Function
1
Mouse or Trackball USB Link
2
Keyboard USB Link
3
Not used
4
Bar Code Reader USB Link
1
2