Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A210 - Gantry cover replacement guide
3-5
Compression Covers
•
Refer to
and
to determine if the defective cover(s)
can be replaced individually, or if a complete set must be ordered and installed.
•
Order and install the specified parts. Use a copy of the table given in
to list the parts to be ordered.
•
After installing the new covers, verify that the positioner reaches rotation and angulation limit posi-
tions correctly, and that no cable is twisted or under excessive strain.
Compression Cover
Upper
Compression Cover