Chapter 3
Page no. 284
Indicators and Switches.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Central Listing
3-1-3.
Mains Distribution Assembly Configuration Jumpers
3-1-4.
Mains Distribution Assembly Fuses
Jumper
Default
Setting
Function
TP1
OFF
(Open)
Control Station Power OFF function.
If TP1 is OFF (as required), the Control Station cannot
•
cannot be powered off by the Generator Control Buttons
•
cannot be powered off by the X-ray Console Power button
•
it is powered off by the ADS shutdown process
TP2
OFF
(Open)
Conditioner auto power off enabled.
If the TP2 jumper is OFF (as required), the Conditioner is always ON even
when the Generator is powered OFF.
TP3
ON
(Closed)
DPS auto power off enabled.
If TP3 is ON (as required), the DPS is switched OFF by system.
Fuse
Specification
PN
Function
F1
250 Vac, 10 A, Delay
99183979
Power Supply (XP4) to Control Station (UPS)
F2
250 Vac, 5 A, Delay
99183923
Power Supply (XP5) to Detector Power Supply input
F4
Line Power Voltage Input to Conditioner (J2 - LED
DS8)
F3
250 Vac, 1.6 A, Delay
99183925
Power Supply to Low Voltage Power Supply input
F5
250 Vac, 1.0 A, Delay
99183936
+ 12V from Low Voltage Power Supply