GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A335 - AWS (U20) Motherboard Battery (V3 Control Stations only)
Page no. 1537
Chapter 9
JC-DR-A-335.fm
Job Card D/R A335 - AWS (U20) Motherboard Battery (V3 Control Stations
only)
Chapter 9
1
SUPPLIES
(5334112)
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4
SAFETY PRECAUTIONS
None
5
PREREQUISITES
None
6
PROCEDURE
1. Remove the X-ray Console.
2. Remove the Control Station top cover. See
Job Card PHY A040 - Remove/Reinstall Control Station
.
3. Remove the Control Station omega cover. See
Job Card PHY A040 - Remove/Reinstall Control Sta-
4. Remove the U20 Workstation from the Control Station (refer to
in
).
5. Turn the power switch on the back of the U20 chassis to the OFF position ( 0 ).
6. Loosen the two captive thumb screws located on the rear edge of the left side access panel.