GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
Page no. 849
Chapter 9
ERR-SUB-A-001.fm
096/024
I06 Abort: Detect
sig. out range
Dose value returned by IDC out of
range
Reboot the full system.
096/025
I08 Track/filter
change
Track/filter configuration changed
between the pre-shot and the exposure
None (warning only).
096/033
I11
IDC sent an invalid dose value to the
Senographe.
Reset the entire system and retry the
exposure.
096/038
I31 Abort: Image
signal too low
Image signal too low.
Occurs in AOP mode, between pre-
exp and exposure, if the dose sent by
IDC is out of permitted range.
Check:
Mechanical thickness.
Paddle type used.
Presence of implant or marking in
the breast.
That paddle is located in the center
of the trolley support.
That the paddle type is not changing
during compression (paddle label
position not correct).
096/039
I32 Abort: Image
signal too high
Image signal too high.
Occurs in AOP mode, between pre-
exp and exposure, if the dose sent by
IDC is out of permitted range.
Check:
Mechanical thickness.
Paddle type used.
Presence of implant or marking in
the breast.
That paddle is located in the center
of the trolley support.
That the paddle type does not
change during compression (paddle
label position not correct).
096/040
I33 Abort No AOP
zone
Exposure aborted; no AOP zone estab-
lished.
Check for presence of breast implant.
096/041
I38 Thermal limit:
mAs not optimal
Due to the system thermal condition,
the exposure mAs value has been
reduced to be within the regular Image
Quality limits
.
None (warning)
096/042
I39 Exp. time limit:
mAs not optimal
Due to exposure time limit, the expo-
sure mAs value has been reduced to be
within the regular Image Quality limits.
None (warning)
096/043
I40 Exp. time limit:
risk mAs too low
Maximum exposure time reached; the
exposure mAs value has been reduced
more than the regular Image Quality
limits.
None (warning)