Chapter 3
Indicators and Switches.fm
The Lift DC Board PL201 and the Rotation DC Board PL202 are different boards located in different
places within the Senographe system. They have similar layout as shown above and indicators as
shown in the table below.
The diodes DS4 and DS8 are associated with lift/rotation movement. For more information, refer to
Card DIAG A010 - No Lift or Rotation Movement
.
LED
Function
Color
Behavior
DS1
15 V
Green
ON when the 15 V supply voltage is OK
OFF when the 15 V supply voltage is defective
DS2
3,3 V
Green
ON when the 3.3 V supply voltage is OK
OFF when the 3.3 V supply voltage is defective
DS3
5 V
Green
ON when the 5 V supply voltage is OK
OFF when the 5 V supply voltage is defective
DS4
48 V Power applied on H-Bridge
Red
ON when 48 V DC is applied on the H-Bridge
OFF when 48 V DC is not applied on the H-Bridge
DS5
Rotation Brake Command
(does not exist for lift board or not used
by lift board)
Green
ON when the rotation brake is inactive (rotation
movement is enabled)
OFF when the rotation brake is active (rotation
movement is not enabled)
DS6
Clutch Command
(does not exist for lift board)
Green
ON when the clutch is inactive (Angulation move-
ment is enabled)
OFF when the clutch is active (Angulation move-
ment is not enabled)
DS7
Microcontroller activity
Green
Blink fast = node not initialized
Blink slow = OK
Light ON = CAN activity
DS8
Verifies whether power is effectively
applied to the motor when a movement
command has been requested.
Red
OFF if a movement command is sent from the MOS
drive and power is effectively applied to the motor.
ON if no movement command is sent from the
MOS drive and no power is effectively applied to
the motor, or if there is a faulty H-Bridge.
DS9
RT2 Grid Synchro
Yellow
DS10
CAN Bus activity
Yellow
ON or OFF when CAN Bus activity occurs.
DS11
Not Used
Yellow
N/A
DS12
RT1 valid Line
Yellow
ON when the valid line is active (movement is
enabled)
OFF when the valid line is inactive (movement is
not enabled)
DS13
DSP Live
Green
Blinks fast when the DSP is alive, representing the
heartbeat of the DSP.
OFF when the DSP is dead.
DS14
3.3 VA
Green
ON when the 3.3 VA analog supply voltage is OK
OFF when the 3.3 VA analog supply voltage is
defective