GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
Page no. 107
Chapter 3
S2100D How To.fm
PRESSING THE "MEASUR" KEY UP-
DATES THE DISPLAYED VALUE OF THE
THEORETICAL MEASURED X-RAY
TUBE CURRENT VALUE
GENERATOR CALIBRATION
BIAS
TUBE
HTR_SCL
mA_meas
KV
kV CALIBRATION
CALIB
PARAM
kV CALIBRATION
1st pt
calcul
2nd pt
kV CALIB 1st point ref=(value) kV
kV CALIB 1st point
kV REFERENCE 1st point ref=(value) kV
CALIB
REF
kV_M
NEXT
VALID
MEASURED kV 1st meas=(value)
NEXT
VALID
Same menu
structure as for
1st pt
VALID
THIS MENU IS ONLY ACCESSIBLE
WHEN BOTH HV VALUES HAVE
BEEN MEASURED AND ENTERED
VIA THE "1ST PT" AND "2ND PT"
MENUS.
kV PARAMETERS x=(2 values)
CHANGE
PARAM kV MODIF x=(value)
NEXT
VALID
mA MEASUREMENT
CALIB
PARAM
mA MEASURE CALIBRATION
CALIB
I_MEAS
CALCUL
mA MEASURE CALIBRATION meas=x mA
MEASUR
MEASURED CURRENT meas=(value)
NEXT
VALID
mA MEASUREMENT CALC
VALID
mA MEASUREMENT PARAM f/mA=(value)
CHANGE
mA MEAS PARAM MODIF f/mA=(value)
NEXT
VALID
THIS MENU IS ONLY
ACCESSIBLE WHEN X-RAY
TUBE CURRENT HAS BEEN
MEASURED AND ENTERED
VIA THE "CALIB" AND
"I_MEAS" MENUS.
B
THIS MENU IS ONLY
ACCESSIBLE WHEN 1ST PT
KV VALUE HAS BEEN
MEASURED VIA THE "CALIB"
MENU.
GENE
BIAS
mA_meas
kV