Chapter 9
Page no. 1140
JC-DR-A-191.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A191 - PDU Board
PDU board cable connections:
9. Ensure that jumper JP22 (5) on the PDU board is open and only connected to the very top connector.
10. Ensure that jumper PRGM (6) on the PDU board is open.
11. Switch on the Senographe electrical power.
Connector
Label
Description of cable
(usually identified by code on a yellow label)
Cable color
J1
J1/PL101-W104
Gray
J2
J2/PL101-W125
White
J3
No code, but cable ends with a gray ribbon cable
Black
J4
2345987
Black
J5
J5/PL101-W110
Gray
J6
J6/PL101-W315
Gray
J7
Three cables arriving at the same physical connector:
J7/PL101-W114
J7/PL101-W209
2345992
Black
Gray
Gray
J8
J8/PL101-W101
Gray
J9
No cable attached
N/A
J10
No cable attached
N/A
5
6