Chapter 9
Page no. 868
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
112/004
N/A
ROTATION
MOTION UNCO-
HERENT
Occurs when pulses are sent by the coder whereas
the IPMS considers that there must not be any
motion. This check concerns the Lift, Rotation, Com-
pression, and Tilt nodes.
Reset the system from the X-ray Console. If the
problem persists call your OLC.
112/005
HALT
ROTATION CORRE-
LATION ERROR
Occurs when signals from absolute coder (potenti-
ometer) and coding wheel (motor encoder) are not
correlated.
Refer to
JC DIAG A1107 ROTATION CONTROL
to fix the problem.
A1107
112/008
HALT
ROTATION BRAKE
DEFECT
Occurs when the Arm rotation brake has no current
or less than 1.5 A running through it. The brake cur-
rent is continuously measured, and this alarm is
raised if the brake:
- is released and there is not enough current through
it
- is operating and there is too much current in it
Refer to
to fix the problem.
A1112
112/010
HALT
ROTATION HARD-
WARE LIMIT
REACHED
Occurs when the IPMS detects that a motion has
reached the electrical stop limit. In this case the
IPMS stops and forbids the motion and sends an
error message.
Refer to
JC DIAG A1109 ROTATION CONTROL
to fix the problem.
A1109
112/011
HALT
ROTATION OVER-
SPEED
Occurs when the speed measured by encoder goes
too high as regard to the speed profile (and correla-
tion is OK).
1. First, refer to
2. If the problem persists, check rotation mecha-
nism, if greasing required, refer to
A205 - Mechanical checks and lubrication
.
A1113
PM A205
112/012
WARNING
ROTATION LOW
SPEED
The rotation speed measured by the encoder is too
low in relation to the speed profile (and correlation is
OK).
Refer to
JC DIAG A1114 ROTATION CONTROL
to fix the problem.
A1114
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card