Job Card CHK A003 - UPS Batteries
10. Wait until the UPS battery is fully recharged (i.e. when the 5 charge LEDs on the UPS are illu-
mintated. Once the UPS battery is fully recharged, restore the parameters using the steps below.
11. Power up the Senographe system, and login as usual. The from behind the Browser, launch a com-
mand window (see
Launching a Command Line Window
).
A command window appears with an ADS prompt.
12. Change to the
/export/home/insite/browser
directory, by entering:
cd /export/home/insite/browser
13. Connect to the UPS web-based diagnostics, by entering:
mozilla-bin http://localhost:3052
14. At the login page that appears, specify
root
for the username and
root
for the password.
15. Use the UPS web-based interface to restore the UPS paramters as follows:
a. Select the left menu
Event
, then
Actions
, and click on the
Battery discharged
event.
b. Under the
Command file
item:
- check the
Enable command file execution
checkbox
- select in list
/opt/APC/PowerChuteBusiness/Agent/cmdfiles/batteryDischarged.sh
- enter a value of 3 seconds for
Command file Execution Duration
c. Under
Shutdown
, check the
Initiate Shutdown Sequence
box.
d. Click on the
Apply
button.
e. Click the
Insufficient Runtime Available
event.
f. Under
Shutdown
, check
Initiate Shutdown Sequence
.
g. Click on the
Apply
button.