GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A316 - SB-150 CD-ROM Drive
Page no. 1459
Chapter 9
JC-DR-A-316.fm
! Notice:
Use proper ESD grounding techniques when handling components. Wear an antistatic wrist strap
and use an antistatic mat. Store ESD-sensitive components in antistatic bags before placing them
on any surface.
10. Remove the IDE cable (1) and power cable (2) from the back of the CD-ROM drive.
11. Remove the two screws (cross-head screwdriver) securing the CD-ROM drive to the chassis.
12. Place your fingers on the back of the CD-ROM drive. Push the CD-ROM DVD-ROM drive toward the
chassis front and remove it.
13. Place the faulty CD-ROM in an antistatic bag.
14. Take the new CD-ROM drive, and verify that the drive back panel mode-select jumper is set to
1
2