Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A016 - Rotation Movement Troubleshooting
2
TROUBLESHOOTING STEERING GUIDE
Use the following flow chart to isolate the issue causing the Rotation movement problems.
ILLUSTRATION 1 - TROUBLESHOOTING FLOWCHART
Perform Rotation Functional Checks.
See
.
NOK
Grease Checks
See
,
Grease Rotation Parts
See
.
NOK
OK
NOK
Rotation Gas Spring Checks
See
,
NOK
Change Rotation Gas Spring
See
OK
NOK
OK
OK
Check Rotation Motor
See
.
NOK
Change Rotation Motor
See
OK
NOK
Contact OLC.
Change Rotation Board
See
.
OK
NOK
OK
Perform Rotation Functional Checks.
See
.
Perform Rotation Functional Checks.
See
,
Perform Rotation Functional Checks.
See
,
Perform Rotation Functional Checks.
See
.
Cable Checks
See
,
.
Re-route cables
See
.
NOK
OK
NOK
OK
Corrective actions
Perform Rotation Functional Checks.
See
,