GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A405 - HV Cable
Page no. 1567
Chapter 9
JC-DR-A-405.fm
6-2-3
Reference Mark M3.
6-2-4
Reference Mark M4.
6-2-5
Reference Mark M5.
7
COMPLETION
Check for correct operation of X-ray exposures.
Reference mark M3
Mark M3 must be positioned at the level of
the bottom surface of the metal clamp,
which secures the cable harness as it exits
the bottom of the arm.
Mark M4 must be positioned at the
external surface of the column.
Reference mark M4
Mark M5 must be positioned at the point
where the cable leaves the cable sheath
at the input to the Generator cabinet.
Reference mark M5