GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Page no. 731
Chapter 7
JC-ELE-A-043.fm
7
18" MONITORS
The table below summarizes the steps required to calibrate an 18" Monitor.
7-1
18" Monitor Keypad Usage
The monitor keypad at the bottom right corner of the monitor has four buttons, whose functionality is
summarized below.
Control
Station
Monitor
Keypad
Unlocking
OSD
Preliminary
Configuration
Display
SMPTE
Move
OSD
Calibration
1
V3/V4
2
V1/V2
Button
Action
Reference in this Job Card
Display main menu or select a sub-menu
MENU
Enter a sub-menu or increase a value or validate changes
UP
Decrease a value
DOWN
Jump to upper menu or exit a menu
SET