GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A141 - Image Receptor/Digital Detector
Page no. 1101
Chapter 9
JC-DR-A-141.fm
6
PROCEDURE
6-1
Disassembly of the Digital Detector
1. Release (two screws) and remove the front connector (1)
(screwdriver).
2. Disconnect the optical fiber cables (2) by extracting the
optical connectors by hand.
CAUTION
Do not apply force to fiber cables, and take care
not to bend them excessively.
3. Be sure that at least one of the two coolant pipes has a
distinctive mark (you will have to reconnect them to the
appropriate connectors); then disconnect the coolant sup-
ply pipes (3) by pulling on the connectors (not the pipes).
CAUTION
Be careful to avoid spillage of coolant, especially
onto or near electrical connections. If any coolant
is spilled, it must be cleaned up immediately.
4. Release the screws (4) securing the ground cables (7 mm
open ended wrench).
5. Remove the three screws (5) securing the Bucky inter-
connection plate (6) to the rear wall plate (7) of the
Detector support (3 mm allen wrench).
6. Release the hex screw (8) securing the ground cable
terminal (4 mm allen wrench).
CAUTION
Two people are required for removal and subse-
quent reinstallation of the Digital Detector. At least two of the detector securing screws
must be correctly fitted when the detector is not supported by hand or a suitable stand.
2
3
1
4
5
6
7
8