Job Card ELE A043 - AWS Monitor Calibration
Page no. 749
Chapter 7
JC-ELE-A-043.fm
/ LED settings
is selected, and press
the
Key2
button
to enter the
OSD / LED settings
menu.
3. From the
OSD / LED settings
menu, keep pressing the
Key1
button until
Vertical Position
is selected,
and press
the
Key2/Key3
button
until the position of the OSD is towards the top of the screen.
4. When you have reached the desired position for the OSD, keep pressing the
Key4
button until
Exit
OSD
appears. Then keep pressing the
Key1
button until
Accept changes
is selected, then press the
Key2
button to exit and save the settings.
5. Once the OSD menu is in a convenient position, proceed as follows:
•
If you have a V3/V4 Control Station, go to
for the 19" monitor.
•
If you have a V1/V2 Control Station, go to
for the 19" monitor.
With monitor connected via DVI (V3/V4 CS)