Job Card ELE A021 - Networking Connection and Configuration
Page no. 629
Chapter 7
JC-ELE-A-021.fm
4. At the
Hostname
screen, type the Hostname to be used for the Acquisition Workstation in the
Host-
name
field, and press
<Return
> when done. Then press the
<F2>
key to continue.
Note:
The Hostname will automatically be used as the DICOM A.E. Title for the workstation. If there is
more than one Senographe system on the hospital site, the names chosen must clearly differenti-
ate between them. For example, by include the unit number or name.
Do not use the following characters in the Hostname:
! $ ” ‘ ’ { } [ ] * @ # ? ^ _ / etc.
ONLY USE ALPHANUMERIC CHARACTERS WITHOUT SPACES.
It is also possible to use the characters - (dash) and . (dot).
Ensure that the Hostname is no more than 12 characters long.
The Hostname must be the same value as the
Station Name
5. At the
IP Address
screen, type the external IP address to be used for the Acquisition Workstation in
the
IP address
field, and press the
<Return>
key when done. Then press the
<F2>
key to continue.
6. At the
Subnets
screen, ensure that
Yes
is selected for the
System part of a subnet
option, and press
the
<F2>
key to continue.
- Hostname - - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
A host name must be .............................
.................................................
Hostname :
- IP Address - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
IP addresses must .............................
.................................................
IP address :
- Subnets - - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
> To make a selection .............................
.................................................
System part of a subnet :
X
Yes
No