GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
Page no. 555
Chapter 7
JC-ELE-A-008.fm
7-2-3
Enable the DICOM-1 Mode and Disable All Other Modes
So the Operator cannot select other display modes, only the 1-DICOM mode must be enabled.
1. Press the
Enter
button to display the
Screen Manager
window.
2. On the
Screen Manager
window that appears, leave the
<Setup>
item selected and press the
Enter
button.
3. On the
Setup
window that appears, leave the
<Mode Preset>
item selected and press the
Enter
but-
ton.
4. On the
Setup - Mode Preset
window that appears, use the
Up/Down
navigation menu buttons to
select each mode, and the
Left/Right
navigation buttons to change the state to On/Off, as follows:
-
1 - DICOM
On
-
2 - Custom
Off
-
3 - CAL
Off
-
4 - Text
Off
5. When the settings for each mode are as described and shown above, double tap the
Enter
button to
store the settings and close the
Setup - Mode Preset
window.