GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A210 - Gantry cover replacement guide
Page no. 1203
Chapter 9
JC-DR-A-210.fm
3-4-1
Arm Covers - non-available covers
•
If the reference of the defective part appears in the following table, it cannot be replaced individually.
A complete set of covers must be ordered and installed:
-
Order a set of covers as listed below:
-
Refer to
Arm Cover Rework Assembly.
-
After installation of the Arm Cover Rework Assembly, install the new covers using the information
given in
Job Card PHY A044 - Remove/Reinstall Gantry Covers
.
3-4-2
Arm Covers - available covers
•
If the reference of the defective part appears in the following table, it can be replaced individually.
The table indicates the necessary part(s) to be ordered.
-
Install the new cover(s) using the information given in
Job Card PHY A044 - Remove/Reinstall
.
3-4-3
Post -installation checks
•
After installing the new covers, verify that the positioner reaches rotation and angulation limit posi-
tions correctly, and that no cable is twisted or under excessive strain.
Reference
Description
2375037
2375037-3
Arm Cover Left
2375038
2375038-3
Arm Cover Right
Reference
Description
Quantity
2375037-4
Arm Cover Left
1
2375038-4
Arm Cover Right
1
2412297
Arm Cover Rework Assy
1
2401145-2
Compression Cover Upper
1
2401147
Compression Cover Lower
1
2375043-2
Head Cover Left
1
2375044-2
Head Cover Right
1
2399858-2
GE Logo D60
1
5130589
Kit of screws and caps for covers
1
2352894-2
Handle assembly
1
If the defective part is:
Order the following parts:
Reference
Description
Quantity
2375037-2
Arm Cover Left
2375037-4
Arm Cover Left
1
2375038-2
Arm Cover Right
2375038-4
Arm Cover Right
1
2375037-4
Arm Cover Left
2375037-4
Arm Cover Left
1
2375038-4
Arm Cover Right
2375038-4
Arm Cover Right
1
Order also:
5130589
Kit of screws and caps for covers
1