Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A002 - Unpacking the Control Station
Page no. 399
Chapter 6
JC-PHY-A-002.fm
4. On the right hand side of the Control Station:
•
remove the three hex bolts (3) that secure the bracket to the Control Station base
•
remove the two hex bolts (4) that secure the bracket to the pallet
•
remove the bracket from between the Control Station base and pallet
•
remove the bolt (5) that secures the rectangular bracket to the pallet, and remove the rectangular
bracket
5. With the ramp in place, guide the Control Station off of the pallet.
6-3
Unpacking
1. Unpack the accessories pallet box containing the monitor, the radiation screen, and the accessories.
2. Examine all items for damage. If there are any signs of damage, refer to
. If there are no signs of damage, set them aside until required for
connection.
3. Collect all packing material for disposal.
5
3
4
4