Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
162/005
INHIBIT
GRID NODE NOT
STARTED
The grid node did not start. At power-up, the grid
node sends an acknowledgment (which can be dif-
ferent from the first heartbeat signal) to Gantry CPU.
If the grid node does not send an acknowledgment,
the Gantry CPU creates this error.
1. Reboot the system from the X-ray Console.
2. If the problem persists then change the Bucky
and check the mechanical thickness.
D/R A255
ELE A015
162/006
INHIBIT
TILT NODE NOT
RESPONDING
Possible causes:
• The Tilt node did not send the heartbeat signal.
Between two motions, the tilt node sends a Heart-
beat signal every second to Gantry CPU. If not, the
Gantry CPU creates this error. During lift motion,
the tilt node sends an actual value message regu-
larly in addition to the heartbeat. If not, the Gantry
CPU creates this error.
• The TP1 connector on the Tilt board is closed.
Proceed as follows:
1. Reboot the system from the X-ray Console.
2. If the problem persists check the error logs. If this
error occurs in conjunction with error 162/003,
there is a possibility that a jumper exists on the
TP1 connector on the Tilt board. In this case you
must remove the jumper from the TP1 connector
on the Tilt board.
3. If the problem persists, contact your OLC.
162/007
INHIBIT
GRID NODE NOT
RESPONDING
Possible causes:
• The grid node did not send the heatbeat signal.
Between two motions, the grid node sends a Heart-
beat signal every second to Gantry CPU. If not, the
Gantry CPU creates this error. During lift motion,
the grid node sends an actual value message regu-
larly in addition to the heartbeat. If not, the Gantry
CPU creates this error.
Proceed as follows:
1. Reboot the system from the X-ray Console. If the
problem persists contact your OLC.
2. If the problem persists, contact your OLC.
162/008 ADVISORY
LCD: A14 Risk of
collision with floor
An active angulation motion was stopped or an
angulation UIF request was rejected because a
potential collision situation was detected by the Gan-
try CPU.
Operator error. No service recommended actions.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action