GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
display function
is set to
4
.
If these values are set not as recommended:
•
change them by selecting it with the
Key1
button and change it to the correct value with the
Key2
button
•
store the changed settings by pressing the
Key4
button until
Accept changes
is selected, then
press the
Key2
button to exit and save the settings.
10. Once the changed settings are saved, go to
for the 19" monitor.
INCORRECT
INCORRECT
CORRECT
CORRECT