Chapter 9
Page no. 864
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
103/002 ADVISORY
LCD: A06 Lift end
position is reached
On every “StartMotion” command, the IPMS checks
whether the software stop limit for the specified
direction is not already reached. If the limit is
reached, the "StartMotion" command is ignored and
the following error is returned:
Lift
“UpperLimitAlreadyReached” if the lift upper stop
limit is active
“LowerLimitAlreadyReached” if the lift lower stop
limit is active
Operator error. No service action is required.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card