Chapter 9
Page no. 832
ERR-SUB-A-001.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
•
The following abbreviations are used in the error code descriptions:
IST: Installation
MNT: Maintenance
VRTX: Real time operating system
•
Software errors. Many of the codes are described as
software errors
(e.g., 002/001; Gene. CPU soft-
ware error). These errors normally never occur. If they do, and especially if they are repeated, con-
tact Service Engineering with as much information as possible. Include a copy of the error codes file
on floppy diskette or CD-R. For more information about copying error logs to floppy diskette or CD-R,
refer to