Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A141 - Image Receptor/Digital Detector
Page no. 1099
Chapter 9
JC-DR-A-141.fm
Job Card D/R A141 - Image Receptor/Digital Detector
Chapter 9
1
SUPPLIES
Digital Detector and its floppy disk / CD-ROM with parameters.
Order the digital detector according to your system configuration and country as summarized in the table
below.
TABLE 1 - DETECTOR FRU ORDERING INFORMATION
! Notice:
When investigating a Detector problem, never attempt to open a defective Detector and never at-
tempt to remove the Detector carbon top cover.
If you open a defective Detector or remove the Detector carbon top cover, the Detector warranty
is void and the repair center will not take the responsibility to repair the Detector.
2
TOOLS
3
REQUIRED EFFORT
Personnel: 2 Field Engineers
Time: 45 minutes
4
SAFETY PRECAUTIONS
WARNING
- Power off the Gantry before disconnecting or working on the Digital Detector.
- If the casing is punctured, wear protective gloves and dust masks when removing the
detector. Send the protective items for disposal along with the defective detector.
- The Digital Detector is cumbersome. Two people are required for its removal and installa-
tion (see section
). Follow standard safety practices for handling large com-
ponents.
System Version
Country
FRU
New system with a Failure On Installation (FOI).
All countries
2383172-5
All existing systems in the field that have
received FMI12075DS or have ADS version
32.05 or later (manufactured from July 2006).
Iberia Region
2383172-5
All countries other than Iberia
Region
2383172-4
All existing systems in the field that have ADS
version 32.04 or earlier (manufactured before
July 2006), and that did not receive
FMI12075DS.
All countries
2383172-3
Note:
If 2383172-3 is no longer available, then apply FMI12075DS and order the FRU corresponding to the
country in which the system is located.
FOI=Failure on Installation