ERR SUB A003 - Gantry Error Messages
Page no. 923
Chapter 9
ERR-SUB-A-003.fm
165/003
INHIBIT
Internal software
error
Received message BCC does not match.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/004
INHIBIT
Internal software
error
RX Buffer FULL.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/005
INHIBIT
Internal software
error
TX Buffer FULL.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/006
INHIBIT
GENERATOR DIS-
CONNECTED
A message from the Gantry CPU can not be sent to
Generator because Generator is not connected or
powered down.
Check cable connections between the Generator
and the Gantry CPU Board.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/007
INHIBIT
Internal software
error
Events count not found.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/008
INHIBIT
Internal software
error
Event number from the list not defined.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/009
DEBUG
ONLY
Internal software
error
Revision description not found in file.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/010
INHIBIT
Internal software
error
Messages count not found.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/011
INHIBIT
Internal software
error
Messages number from the list not defined.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/014
INHIBIT
Internal software
error
Message count not defined.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card