GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A331 - UPS
Page no. 1525
Chapter 9
JC-DR-A-331.fm
5. Remove the defective UPS unit from the Control Station by lifting it vertically from the framework and
pulling it towards you.
7-2
Reassembly of New UPS Unit on V3/V4 Control Stations
1. Put the new UPS unit (1) inside the Control Station.
2. Attach the velcro straps (2) to secure the UPS to the Control Station.
1
2