Chapter 7
Page no. 738
JC-ELE-A-043.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
5. Press the
UP
button to display the
User Settings
menu. Keep pressing the
MENU
button until
Reset
User settings
is selected, and press the
UP
button to execute
Reset User settings
.
The AWS monitor restarts.
) and move the OSD to a convenient position (see
necessary.
7. Press the
MENU
button to access the OSD main menu, then keep pressing the
MENU
button until
Brightness/Contrast
is selected, and press
the
UP
button
to enter the
RGB
menu.
The
RGB
menu
appears.
8. Do not change the default Brightness and Contrast values that appear in the RGB menu. The default
Brightness and Contrast values vary according to the monitor release as follows:
•
50 for the Brightness and 50 for the Contrast
•
37 for the Brightness and 24 for the Contrast