ERR SUB A003 - Gantry Error Messages
112/030
WARNING
ROTATION DECEL-
ERATION ERROR
The rotation motion began to slow down but did not
stop before the allowed time.
Refer to
JC DIAG A1115 ROTATION POSITION
OVERSHOOT/ROTATION CONTROL FAILURE
to fix the problem.
A1115
112/031
HALT
ROTATION
MOTION STARTED
IN REVERSE
DIRECTION
The rotation motion was about to go in the reverse
direction to the direction that was requested.
Change the Rotation board then perform a Rotation
Calibration.
D/R A221
CAL 021
112/032
HALT
ROTATION DSP
TIMEOUT
The DSP did not answer to the MCU before the time-
out.
Change the Rotation board then perform a Rotation
Calibration.
D/R A221
CAL 021
112/033
WARNING
ROTATION CAN
COMMAND NOT
PROCESSED
The CAN command was not processed. The main
CPU have sent command to the rotation node before
the rotation node could acknowledge the previous
command.
Reboot the system from the X-ray Console.
Retry the command.
If the problem persists, contact your OLC.
112/034
HALT
ROTATION RX CAN
BUFFER OVERRUN
Possible causes:
• CAN overrun. The input queue of the incoming
CAN messages of the MCU is full and message
dropped.
• The TP1 connector on the Rotation board is
closed.
Proceed as follows:
Check whether a jumper exists on the TP1 connector
on the Rotation board. In this case you must remove
the jumper from the TP1 connector on the Rotation
board.
If there is no jumper on the TP1 connector on the
Rotation board, reboot the complete system. If the
problem persists, check the other nodes (other than
rotation). If the other nodes are OK, check the rota-
tion node (cabling, potentiometer, board).
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card