Job Card D/R A411 - Mains Distribution Board 200-PL1
6. Remove all of the cables from the Mains Distribution board (8 mm open ended wrench). Ensure that
you do not move them too much so they are to re-connect later on.
7. Remove the four bolts (1) (7 mm open ended wrench) which attach the Mains Distribution board to
the Generator chassis.
8. Remove the faulty Mains Distribution board from the Generator chassis.
9. Put the defective Mains Distribution board in an anti-static bag.