GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A001 - AWS Troubleshooting - No Display on Power Up
Page no. 1011
Chapter 9
JC-TSG-A-001.fm
1-2
AWS Power Checks
If the problem persists, start
again on
, or ask for
help from your OLC.
No
Correct the cable problem.
From
UPS and MDR check is good, but
AWS does not power up,
It is likely that the ADS computer PSU is faulty. Plug the ADS computer into a wall
socket. If the ADS computer still does not power up and function correctly then the
ADS computer PSU is faulty.
Is the ADS power
supply faulty?
Yes
Change the ADS Computer
No
Correct the cable problem.
No
Is the UPS-AWS
power cable
incorrectly connected
or defective?
Is the MDR-UPS
power cable
incorrectly connected
or defective?
Yes
Yes