Chapter 11
Page no. 1808
JC-PM-A-429.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A429 - Conditioner Air Filter Cleaning
2. Remove the air filter from its housing. Pull the plastic tab (3)
towards the rear of the unit then pull the tab out to begin dislodg-
ing the air filter (4).
3.
4. Wash the air filter by holding the clean side of the air filter under
running water for either two minutes or until the dirty side of the
filter is clean.
5. Allow the filter to dry.
6. Use a piece of adhesive tape to remove any lint in the filter.
7. Replace the air filter in its housing.
8. Reassemble the air duct and secure it by tightening the screw.
9. Reassemble the Generator Cabinet covers.
3
4