Job Card TSG A015 - Lift Movement Troubleshooting
2-3
TROUBLESHOOTING STEERING GUIDE FOR TYPE 3 LIFT SCREW ASSEMBLY
Use the following flow charts (illustration 6 and illustration 7) to isolate the issue causing the Lift move-
ment problems on systems with a Type 3 Lift Screw Assembly.
ILLUSTRATION 6 - LIFT TROUBLESHOOTING ENTRY - TYPE 3 LIFT SCREW ASSEMBLY
Does customer complain about
Error 102/024 or does Error 102/024 repeatly
appear in the error logs?
Perform the lift trouble shooting steps described
in illustration 7,
Lift Movement Problems Related to Errors 102/
013 and 102/014 - Type 3 Lift Screw Assembly
on
Proceed to the Grease Checks in illustration 7,
and follow the flowchart as required.
NOK
Perform Lift Functional Checks.
See
,
checks, return back here.
OK
No
Yes
Replace Brake Motor
(5413527)
See
.
OK